Tested Opening |
Safety height of 200 mm (8 inch)
|
Maximum Opening |
420 mm (17 inch)
|
Inflow Velocity |
0.53 ± 0.025 m/s
|
Down Flow Velocity |
0.33 ± 0.025 m/s
|
ULPA Filter |
Two, 99.999 % efficient at 0.12 um, filter life indicator
|
Front Window |
Motorized, two-layer laminated toughened glass ≥ 5 mm, anti UV
|
Noise |
EN12469 ≤ 58 dB / NSF49 ≤ 61 dB
|
UV Lamp |
30 W×1 UV timer, UV life indicator, emission of 253.7 nanometres for most efficient decontamination
|
Illuminating Lamp |
12 W×2 LED Lamp
|
Illumination |
≥ 1000 Lux
|
Consumption |
800 W
|
Waterproof Sockets |
Two, total load of two sockets: 500 W
|
Display |
LCD display: exhaust filter and down flow filter pressure, filter and UV lamp working time, inflow and down flow velocity, filter life, humidity and temperature, system working time etc.
|
Control System |
Microprocessor
|
Airflow System |
70 % air recirculation, 30 % air exhaust
|
Visual and audio alarm |
Filter replacement, front window at unsafe height, abnormal airflow velocity
|
Material |
Work Zone: 304 stainless steel;
Main Body: Cold-rolled steel with anti-bacteria powder coating
|
Work Surface Height |
750 mm
|
Caster |
Foot master caster
|
Power Supply |
AC220V±10 %, 50/60 hz; 110V ± 10 %, 60 hz
|
Internal Size (W×D×H) |
1150×600×660 mm
|
External Size (W×D×H) |
1300×750×2250 mm
|
Package Size (W×D×H) |
1460×1050×1800 mm
|
Gross Weight |
280 kg
|
-
Managed and monitored by an integrated microprocessor controller
-
Large LCD digital display for ease of monitoring all safety parameters
-
Incorporated with memory function in case of power-failure
-
Work zone made up of 304 stainless steel
-
Built in with time reserve function, automatic air speed adjustable with filter block and interlock function
-
Integrated with visual and audible alarm for abnormal airflow velocity, filter replacement, front window at unsafe height
-
Designed with universal caster with brake and leveling feet
-
Easy to use, operate and maintain
-
Easy to clean, corrosion resistant, stainless steel lining
Biosafety Cabinet Class II A2 is widely used in pathology, research laboratories, universities, colleges and hospitals. It is suitable to use for microbiology research in absence of volatile, toxic chemicals and radionuclide.