Exterior Size (W × D × H) |
1540 × 800 × 2200 mm
|
|
Interior Size (W × D × H) |
1320 × 670 × 730 mm
|
|
Maximum Opening |
520 mm
|
|
LED Lamp |
16 W × 1
|
|
UV Lamp |
Emission of 253.7 nm for efficient decontamination
|
|
Air Velocity |
0.3 m/s to 0.8 m/s
|
|
Work Surface Height |
850 mm
|
|
Exhaust Duct |
PVC 4 meters, φ 300 mm
|
|
Noise |
≤ 68 dB
|
|
Power Supply |
AC 220 V ± 10%, 50/ 60 Hz
AC 110 V ± 10%, 60 Hz
|
|
Power Consumption |
500 W
|
|
Package Size (W × D × H) |
1690 × 990 × 1690 mm
|
|
Gross Weight |
350 kg
|
-
Microprocessor control system
-
Digital LED display
-
LED lamp for illumination
-
UV lamp for decontamination of working environment
-
9 level adjustable air speed exhaust blower
-
Air velocity memory function in case of power failure
-
Built-in centrifugal blower
-
5 mm Motorized front glass window
-
Active carbon filter to remove toxic fumes
-
Alarm function on filter reaching working time of 3500 hours
-
Chemically resistant phenolic resin work table
-
Interior material highly authentic melamine board resistant to weak acid and base
-
External material made up of cold rolled steel having anti-bacterial coating
-
Backside air compensation avoiding turbulence in working area
Ducted Fumehood is used for providing effective working environment for laboratory experiments in various fields of microbiology, chemical, biotechnology, research centers etc.