Chromatographic Pump |
|
Type |
High-pressure and low –pulse double plunger tandem advection pump |
|
Maximum Pressure |
35 MPa (PEEK) |
|
Flow Range |
0.001 to 9.999 mL/min |
|
Flow accuracy |
± 0.5 % |
|
Flow Repeatability |
RSD ≤ 0.1% |
|
Flow stability |
(0.2 to 0.5) mL/min, ≤ 3 %
(0.5 to 1.0) mL/min, ≤ 2 %
>1.0 mL/min, ≤ 2 % |
|
|
Built-in Heat conduction type column heater |
|
Temperature range |
Room temperature +5°C to 60°C (41 to 140°F) |
|
Allowable deviation |
± 1°C |
|
Temperature stability |
≤ 0.5°C/h |
|
|
Electromagnetic injection valve |
|
Maximum pressure |
35 MPa |
|
Control mode |
By stepper motor |
|
Power supply |
24 V (DC) |
|
|
Digital and temperature control detection system |
|
Type |
Temperature control and bipolar conductivity detector |
|
Detection type |
Bipolar conductivity detection |
|
Cell volume |
≤ 0.8 µL |
|
Measurement range |
0 to 45000 µ S/cm (adjustable) |
|
Detection resolution |
≤ 0.0020 nS/cm |
|
Output voltage |
-6000 to 6000 mV (adjustable) |
|
Temperature range |
Room temperature +5°C to 60°C (41 to 140°F) |
|
Temperature accuracy |
± 0.01°C |
|
Maximum pressure |
10 MPa |
|
Baseline drift |
≤ 20 % FS/30 min |
|
Baseline noise |
≤0.5 % FS |
|
Instrument linearity |
≥ 0.999 |
|
Quantitative Repeatability |
≤0.5 % |
|
Qualitative Repeatability |
≤2 % |
|
Minimum detectable concentration |
Cl-≤0.005 µg/mL
Li+ ≤0.005 µg/mL |
|
|
Flow system |
|
Plastic path flow |
Made of PEEK material |
|
Six way valve |
PEEK material pressure 5000 psi, independent automatic collecting and flow function |
|
|
Panel computer |
|
Display screen |
12.3 inch |
|
Internal memory |
2G |
|
Weight |
786 g |
|
|
External size (L × W × H) |
440 × 226 × 420 mm
|
|
Net weight |
8 kg
|
|
Gross weight |
11 kg
|
-
Built-in column temperature control box for accurate experimental results
-
Capable with Flash column for rapid detection of anions and cations
-
Intelligent flow path clearing system for easy cleaning
-
Real time display
-
Wi-Fi connection for convenient data transfer
-
Continuous, non-stop 10 hours working system
-
Easy to operate
Used for on field selective analysis of organic and inorganic ions, also used in quality and research laboratories across pharmaceutical, water quality, chemical industries.