Our LOFP-B10 is deliberately furnished to determine both flash point and fire point. It has a wide range of applications with a great market value. It is designed as per the standard ASTM D92, ISO2592, IP36 (Standard Test Method for Flash Point of Petroleum Products-Cleveland Open Cup Method).
Get Catalog| Ambient Temperature | 10 ℃ ~ 40 ℃ |
| Relative Humidity | ≤ 85 % |
| Temperature Measurement- Full Scale | Ambient temperature to ~ 400 ℃ |
| Temperature Measurement- Repeatability | ≤ 4 ℃ |
| Test accuracy | 0.005 |
| Reappearance | ≤ 8 times |
| Graduation | 0.1 ℃ |
| Dimensions | 620 x 500 x 440 mm |
| Power Consumption | 400 W |
| Power Supply | AC 220 V ± 10 %, 50 Hz |
It can be widely used in railway, aviation, electric power, petroleum industry and colleges, scientific research institutes, measurement departments to do determination of flash point and fire point for petroleum products except oil fuel and samples whose open cup flash point is below 79 ℃
Request Quote

| Measuring Range For Aviation Gasoline | 86.0 to 105.0/MON |
| Measuring Range For Gasoline For Motor Vehicles | National standard: 65.0 to 110.0/RON National-V standard: 70 to 115.0/RON Blended gasoline: 75.0 to 120.0/RON Light gasoline: 50.0 to 85.0 /MON Isomerized gasoline: 85.0 to 120.0/RON Arene gasoline: 90.0 to 120.0/RON |
| Measuring Range For Ethanol | E93: 92.0 to 95.0/RON E97: 95.0 to 99.0/RON Judgement: 60.0 to 120.0/RON |

| Measurement range | 0 to 600 penetration scale |
| Timing controller | 5 sec, 8 sec, 10 sec, 12 sec, 30 sec and 60 sec |
| Time accuracy | ± 0.1 seconds |

| Measurement range | 0 to 600 penetration scale |
| Timing controller | 5 sec or 60 sec |
| Time accuracy | ± 0.1 seconds |
Copyright @ labtrononline.com