Automatic Dielectric Tester LADT-A10 is a fully automatic oil tester. The oil tester adopts industrial SCM to control which is applied to integrate circuit. It features anti-jamming technology which largely improves the machine capacity.
Get Catalog| Capacity | 2.0 kVA |
| Speed of voltage increase | 2 k V/ s |
| Voltage inspection accuracy | 0.01 |
| Breakdown sensitivity | < 0.1 kV |
| Waveform distortion | ≤ 1 % |
| Breakdown response time | ≤ 10 MS |
| Polar distance | 2.5 mm |
| Storage environment |
Temperature: 0 ℃ -40 ℃ Humidity: Max. relative humidity 85% |
| Power supply |
Temperature: - 20 ℃ - 60 ℃ Humidity: Max. relative humidity 75% |
| Output voltage | AC 220 V ± 10 %, 50 Hz / 60 Hz |
| Dimension | 60 kV |
| Weight | 380mm × 380mm × 380mm |
Used to test the insulating oil’s dielectric strength
Request Quote

| Measuring Range For Aviation Gasoline | 86.0 to 105.0/MON |
| Measuring Range For Gasoline For Motor Vehicles | National standard: 65.0 to 110.0/RON National-V standard: 70 to 115.0/RON Blended gasoline: 75.0 to 120.0/RON Light gasoline: 50.0 to 85.0 /MON Isomerized gasoline: 85.0 to 120.0/RON Arene gasoline: 90.0 to 120.0/RON |
| Measuring Range For Ethanol | E93: 92.0 to 95.0/RON E97: 95.0 to 99.0/RON Judgement: 60.0 to 120.0/RON |

| Measurement range | 0 to 600 penetration scale |
| Timing controller | 5 sec, 8 sec, 10 sec, 12 sec, 30 sec and 60 sec |
| Time accuracy | ± 0.1 seconds |

| Measurement range | 0 to 600 penetration scale |
| Timing controller | 5 sec or 60 sec |
| Time accuracy | ± 0.1 seconds |