HPLC system LHLC-B10

This HPLC system is incorporated with advanced high pressure injection technique. The detector is equipped with 24 bits AD transducer to assure low noise, low drift and high sensibility. This HPLC system can conveniently change the sample input volume. It comes with dual piston reciprocating and series pump for consistent and stable flow.

Get Catalog
High pressure pump High pressure gradient pump
Peak operating pressure 45 MPa, Stainless steel pump
Pressure ± 0.02 MPa
Flow rate range 0.001 - 9.999 mL / min
Flow rate precision ± 5 %
Flow rate repeatability RSD ≤ 0.1 %
Power 220V ± 10 %, 50 Hz, 250 W
Size (W × H × D) 520 × 420 × 160 mm
UV Detector
Wavelength 190 - 680 nm
Wavelength error ≤ ± 1 nm
Baseline noise ≤ ± 2 × 10-5 AU (Methyl alcohol, 1 mL / min, 254 nm, 20 ºC)
Wavelength repeatability 0.2 nm
Flow cell 8 µL
Drift ≤ 5 × 10-4 AU/h (Methyl alcohol, 1 mL / min, 254 nm, 20 ºC)
Limit of detection ≤ 2 × 10-9 g / mL (Naphthalene methyl alcohol solvent)
Automatic Injector
Sample value Maximum 2 mL, 6 × 25 = 150 units
Sample volume 0 - 1mL, adjustable
Sample residual < 0.02 %
Column oven
Temperature control Room temperature ± 5 ºC - 80 ºC
Constant temperature accuracy ± 1 ºC
Constant temperature precision ≤ 0.1 ºC
Temperature setting resolution ratio 0.1 ºC
Overall Instrument
Quantitative analysis RDS 6 ≤ 0.2 %
Quantitative repeatability RDS 6 ≤ 0.2 %
Sample residual ≤ 0.01 %
Gradient error ≤ 0.01 %
  • Inclusion of high performance patent gradient mixer
  • Intelligent anti-control software with powerful control function
  • Stand-alone TFT colour screen with touch control
  • High pressure gradient pump

This gradient HPLC system has multiple applications in medicine analysis, detection of protein content of food, environmental analysis, oil and chemical analysis and in scientific research studies.

Request Quote

Standard HPLC

Standard HPLC : HPLC system LHLC-A10
High pressure pumpSingle pump
Flow rate range0.001 - 9.999 mL / min (0.001 mL / min increment)
Pressure range≤ 0.5 MPa (0 - 42 MPa)
Standard HPLC : HPLC system LHLC-A11
High pressure pumpTwo sets of pump
Flow rate range0.001 - 9.999 mL / min (0.001 mL / min increments)
Pressure range≤ 0.5 MPa (0 - 42 MPa)
Standard HPLC : HPLC system LHLC-B11
High pressure pumpHigh pressure constant-current infusion pump
Infusion WayMicro volume series dual-piston
Infusion pressure0 - 9999 Psi

Analytical HPLC

Analytical HPLC : Analytical HPLC System LAHS-A10
Pump specificationQuaternary Low Pressure Pump, 0.001 to 10.000 ml/min, in 0.001ml increments, ≤ ± 0.2% (1.00 ml/min, H2 O, R.T., 8.5 MPa), RSD ≤ 0.06% (1.00 ml/min, H2 O, R.T., 8.5MPa), 63 MPa, ≤ ± 3.0 %, ≤ 0.1 MPa, 0 to 100 %, ≤ 0.3 % RSD
UV/VIS Detector190 to 800 nm, ≤ 0.1 nm, ≤ ± 1.0 nm, 6 nm, 0.0 s to 5 s, ≤ ± 0.25 × 10-5 AU(empty cell, 254 nm, 1.0s), ≤ 0.5 × 10-4 AU/h (empty cell, 254 nm, 60 min), 10 mm(analytical cell), Deuterium Lamp and Tungsten Lamp
Data-processing WorkstationSystem contain the tables for regular samples Equipped with Internal and external standard calculation process Automatic display, print, export of data and starting of another program on completion on measurement Observe with the condition of 21 CFR part 11 directive of the FDA
Analytical HPLC : Analytical HPLC System LAHS-B10
High Pressure Constant Flow Pump1 set, 0.001 to 10.00 mL/min, 0.001 mL/min, ≤ ± 0.3 %, RSD ≤ 00.1 %, 42 MPa, ≤ 1.0 %, ≤ 1.0 %, 80 W
UV/Visible Detector190 to 700 nm, ± 0.1 nm, 8 nm, 0 to 4.9 seconds, 1.8 AU (5 %), ≤ ± 1 × 10-5AU, ≤ 2.0 × 10-4AU/h, 4.0 × 10-9 g/mL, Optical path: 10 mm Volume: 10 µL, Deuterium lamp, 100 W, AC 110 V/ 220 V ± 10 %, 50/60 Hz, 600 × 440 × 380 mm 2 × (600 × 410 × 300 mm, 54 kg
Analytical HPLC : Analytical HPLC System LAHS-B11
High Pressure Constant Flow Pump2 sets, 0.001 to 10.00 mL/min, 0.001 mL/min, ≤ ± 0.3 %, RSD ≤ 00.1 %, 42 MPa, ≤ 1.0 %, ≤ 1.0 %, 80 W
UV/Visible Detector190 to 700 nm, ± 0.1 nm, 8 nm, 0 to 4.9 seconds, 1.8 AU (5 %), ≤ ± 1 × 10-5AU, ≤ 2.0 × 10-4AU/h, 4.0 × 10-9 g/mL, Optical path: 10 mm Volume: 10 µL, Deuterium lamp, 100 W, AC 110 V/ 220 V ± 10 %, 50/60 Hz, 600 × 440 × 380 mm 3 × (600 × 410 × 300 mm, 75 kg